Exchanging the Wash Seals
When
When maintaining seal wash option
Tools required
• 4-mm hexagonal key
Parts required
#
1
6
Preparations
• Switch off pump at the main power switch
• Remove the front cover
1
Disconnect all capillaries and tubes from the pump head
and disconnect the active inlet valve cable.
1200 Series Quarternary Pump User Manual
Part number
Description
0905-1175
Wash seal
5062-2484
Gasket, seal wash (pack of 6)
2
Using a 4-mm hexagonal key stepwise loosen and remove
the two pump head screws and remove the pump head
from the pump drive.
7
Maintenance
Simple Repairs