Remote Cables - Agilent Technologies 1200 Series Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Remote Cables

One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent 1100/1200
Series modules. The other end depends on the instrument to be connected to.
Agilent 1100/1200 to 3390 Integrators
Connector 01046-60203
1200 Series Quarternary Pump User Manual
Pin 3390
Pin Agilent
1100/1200
2
1 - White
NC
2 - Brown
7
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
NC
7 - Red
NC
8 - Green
NC
9 - Black
Identifying Cables
Remote Cables
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
9
125

Advertisement

Table of Contents
loading

Table of Contents