Exchanging the Purge Valve Frit or the Purge Valve
When
Frit – when plunger seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 ml/min of H
Purge valve – if internally leaking
Tools required
• Wrench 1/4 inch
• Wrench 14 mm
• Pair of tweezers or toothpick
Parts required
#
5
1
Preparations
• Switch off pump at the main power switch
• Remove the front cover
1 Using a 1/4 inch wrench disconnect the pump outlet capillary at the purge
2 Disconnect the waste tube. Beware of leaking solvents due to hydrostatic
3 Using the 14 mm wrench unscrew the purge valve and remove it.
4 Remove the plastic cap with the gold seal from the purge valve.
1200 Series Quarternary Pump User Manual
Part number
Description
01018-22707
PTFE frit (pack of 5)
G1311-60009
Purge valve
valve.
pressure.
Simple Repairs
0 with purge valve opened)
2
7
Maintenance
85