Exchanging the Active Inlet Valve Cartridge

When
If internally leaking (backflow)
Tools required
• Wrench 14 mm
• Pair of Tweezers
Parts required
#
1
1
Preparations
Switch off pump at the main power switch and unplug the power cable
Figure 14
1 Remove the front cover.
2 Unplug the active inlet valve cable from the connector.
1200 Series Quarternary Pump User Manual
Part number
Description
G1312-60025
Active inlet valve body
5062-8562
Valve cartridge (400 bar)
Exchanging the Active Inlet Valve
7
Maintenance
Simple Repairs