Early Maintenance Feedback (Emf); Emf Counters - Agilent Technologies 1200 Series Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Early Maintenance Feedback (EMF)

Maintenance requires the exchange of components in the flow path which are
subject to mechanical wear or stress. Ideally, the frequency at which
components are exchanged should be based on the intensity of usage of the
instrument and the analytical conditions, and not on a predefined time
interval. The early maintenance feedback (EMF) feature monitors the usage of
specific components in the instrument, and provides feedback when the
user-settable limits have been exceeded. The visual feedback in the user
interface provides an indication that maintenance procedures should be
scheduled.

EMF Counters

The pump provides a series of EMF counters for the pump head. Each counter
increments with pump use, and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded. Each
counter can be reset to zero after maintenance has been done. The pump
provides the following EMF counters:
• Pump Liquimeter
• Pump seal wear
Pump Liquimeter
The pump liquimeter displays the total volume of solvent pumped by the pump
head since the last reset of the counters. The pump liquimeter can be assigned
an EMF (maximum) limit. When the limit is exceeded, the EMF flag in the user
interface is displayed.
Seal Wear Counters
The seal wear counters display a value derived from pressure and flow (both
contribute to seal wear). The values increment with pump usage until the
counters are reset after seal maintenance. Both seal wear counters can be
assigned an EMF (maximum) limit. When the limit is exceeded, the EMF flag in
the user interface is displayed.
1200 Series Quarternary Pump User Manual

Early Maintenance Feedback (EMF)

7