Exchanging the Optional Interface Board

When
Board defective
Parts required
#
1
Electrostatic discharge at electronic boards and components
C A U T I O N
Electronic boards and components are sensitive to electrostatic discharge (ESD).
1 Switch off the pump at the main power switch, unplug the pump from line
2 Disconnect cables from the interfaceboard connectors.
3 Loosen the screws. Slide out the interface board from the pump.
4 Install the new interface board. Secure screws.
5 Reconnect the cables to the board connector.
6 Reconnect the pump to line power.
1200 Series Quarternary Pump User Manual
Part number
Description
G1351-68701
?
In order to prevent damage always use an ESD protection (for example, the ESD
wrist strap from the accessory kit) when handling electronic boards and
components.
power.