START and STOP are connected via diodes to pin 3 of the 3394 connector.
N O T E
Agilent 1100/1200 to 3396A Integrators
Connector03394-60600
Agilent 1100/1200 to 3396 Series II / 3395A Integrators
Use the cable
Otherwise the integrator prints START; not ready.
1200 Series Quarternary Pump User Manual
Pin 3394
9
NC
3
NC
NC
NC
5,14
1
NC
13, 15
part number: 03394-60600
Pin Agilent
Signal Name
1100/1200
1 - White
Digital ground
2 - Brown
Prepare run
3 - Gray
Start
4 - Blue
Shut down
5 - Pink
Not connected
6 - Yellow
Power on
7 - Red
Ready
8 - Green
Stop
9 - Black
Start request
Not connected
and cut pin #5 on the integrator side.
9
Identifying Cables
Remote Cables
Active
(TTL)
Low
Low
Low
High
High
Low
Low
127