9
Identifying Cables
Remote Cables
Agilent 1100/1200 to 3392/3 Integrators
Connector01046-60206
Agilent 1100/1200 to 3394 Integrators
Connector01046-60210
126
Pin 3392/3
Pin Agilent
1100/1200
3
1 - White
NC
2 - Brown
11
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
9
7 - Red
1
8 - Green
NC
9 - Black
Pin 3394
Pin Agilent
1100/1200
9
1 - White
NC
2 - Brown
3
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
5,14
7 - Red
6
8 - Green
1
9 - Black
13, 15
1200 Series Quarternary Pump User Manual
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Not connected