3
Installing the Pump
Flow Connections of the Quaternary Pump
2 Place the vacuum degasser and the solvent cabinet on top of the quaternary
3 Put the bottle-head assemblies into solvent reservoirs containing your
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
5 Connect the solvent tubes to the outlet connectors (typically right
6 Connect the syringe adapter from the degasser accessory kit to the solvent
7 Using a piece of sanding paper connect the waste tubing to the purge valve
8 If the pump is not part of a Agilent 1200 Series System stack or placed on
9 Connect the pump outlet capillary (pump to injection device) to the outlet
42
pump.
mobile phase and place the bottle in the solvent cabinet.
connectors A to D (typically the left connection of the channel) of the
vacuum degasser. Fix the tubes in the tube clips of the vacuum degasser.
connection of the channel) of the vacuum degasser.
tube of channel A.
and place it into your waste system.
the bottom of a stack, connect the corrugated waste tube to the waste outlet
of the pump leak handling system.
of the purge valve.
1200 Series Quarternary Pump User Manual