Overview Of The Pump's Indicators And Test Functions - Agilent Technologies 1200 Series Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Overview of the Pump's Indicators and Test Functions

Status Indicators
The quaternary pump is provided with two status indicators which indicate
the operational state (prerun, run, and error states) of the quaternary pump.
The status indicators provide a quick visual check of the operation of the
quaternary pump (see
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the instrument
generates an error message in the user interface. For details on error
messages and error handling, please refer to the Agilent Lab Monitor &
Diagnostic Software.
Pressure Test
The pressure test is a quick test designed to determine the pressure tightness
of the system.
Leak Test
The leak test is a diagnostic test designed to determine the pressure tightness
of the quaternary pump.
1200 Series Quarternary Pump User Manual
Troubleshooting and Test Functions

Overview of the Pump's Indicators and Test Functions

"Status Indicators"
on page 66).
6
65