Solvent Information
Always filter solvents through 0.4 µm filters, small particles can permanently
block the capillaries and valves. Avoid the use of the following steel-corrosive
solvents:
• Solutions of alkali halides and their respective acids (for example, lithium
• High concentrations of inorganic acids like sulfuric and nitric acid,
• Halogenated solvents or mixtures which form radicals and/or acids, for
• Chromatographic grade ethers, which can contain peroxides (for example,
• Mixtures of carbon tetrachloride with 2-propanol or THF dissolve stainless
1200 Series Quarternary Pump User Manual
iodide, potassium chloride, and so on).
especially at higher temperatures (replace, if your chromatography method
allows, by phosphoric acid or phosphate buffer which are less corrosive
against stainless steel).
example:
→
2CHCl
+ O
2COCl
3
2
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the stabilizing
alcohol.
THF, dioxane, di-isopropylether). Such ethers should be filtered through
dry aluminium oxide which adsorbs the peroxides.
steel.
+ 2HCl
2
Using the Quaternary Pump
Solvent Information