7
Maintenance
Simple Repairs
11 Place the support rings on the plunger housing (plungers not installed) and
12 Insert the plungers and carefully press them into the seal.
13 Tighten the lock screw.
14 Slide the pump head assembly onto the metering drive. Apply a small
15 Reconnect all capillaries, tubes and the active inlet valve cable to its
16 Route the wash inlet tube into a bottle filled with a mixture of distilled
17 Route the outlet of the wash tube into a waste container.
96
snap the pump head and plunger housing together.
amount of pump head grease (part number 79846-65501) to the pumphead
screws and the balls of the spindle drive. Tighten the pumphead screws
stepwise with increasing torque
connector.
water and isopropanol (90/10) and place the bottle above the pump in the
solvent cabinet.
Figure 22
Pumphead after completed installation.