Removing The Pump Head Assembly - Agilent Technologies 1200 Series Manual

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

7
Maintenance
Simple Repairs

Removing the Pump Head Assembly

When
• Exchanging the seals
• Exchanging the plungers
• Exchanging seals of the seal wash option
Tools required
• Wrench 1/4 inch
• 4-mm hexagonal key
Preparations
Switch off pump at the main power switch and unplug the power cable
Damage of the pump drive
C A U T I O N
Starting the pump when the pump head is removed may damage the pump drive.
1 Remove the front cover.
2 Disconnect the Active Inlet Valve cable.
3 Using a 1/4 inch wrench remove the outlet capillary.
4 Disconnect the capillary from the Outlet Ball Valve.
5 Remove the waste tubing and disconnect the Active Inlet Valve tubing.
6 Remove the capillary at the bottom of the Pumphead.
7 Using a 4 mm hexagonal key, stepwise loosen the two Pumphead screws and
88
Never start the pump when the pump head is removed.
remove the Pumphead from the Pump Drive.
1200 Series Quarternary Pump User Manual