Description:
The interval time is a system relative entry which can be specified in hours and minutes. It is the time
interval between each automatic prompt for a PV Test. For example: The detector is to have a PV Test
every hour therefore the interval time is set to 01.00 which is equal to an hourly prompt. The interval
time cannot be set below the Test Window time plus 1 minute.
Press the [ENTER] key, and then the [s] and [t] to set the required time.
Scope:
Test Window time + 1 minute to 24.00
Default:
01.00 (1 hour)
Notes:
This entry only appears if PV Test is enabled in the OPTIONS menu, and Interval timing has been
selected in the PV Test setup menu.
To ensure PV Test prompts occur at the precise interval time set, irrespective of the time to
successfully complete a PV Test then synchronised timing must be used.