Compensation - Loma Superscan Micro ISC Manual

Hide thumbs Also See for Superscan Micro ISC:
Table of Contents

Advertisement

MENUS A to Z
L o m a S u p e r s c a n M i c r o
CANCEL
CHANGE
Description:
This value is automatically derived during a calibration cycle, and should therefore not normally
require adjustment: only alter the value if advised to do so by a LOMA Service Engineer.
Press the [ENTER] key to view the compensation and signal level values; the [ ] and [ ] keys to
adjust the compensation value. The compensation value is displayed on the left hand side of the
display, and the signal level on the right hand side.
Scope:
-65535 to 65535
Notes:
This menu position is only accessible if the Manual Setup menu has been enabled in the User Options
menu.
The calibrate LED will illuminate whilst the compensation value is being adjusted.
For more information about what the compensation value does, please refer to the Principles of
Operation section.