False Threshold - Loma Superscan Micro ISC Manual

Hide thumbs Also See for Superscan Micro ISC:
Table of Contents

Advertisement

MENUS A to Z
L o m a S u p e r s c a n M i c r o
CANCEL
CHANGE
Description:
The false threshold is used to identify metal test samples being used which are greater than those
specified for testing.
A product relative false threshold default value is generated for each product position. The value can
then be manually adjusted to achieve precise control.
Any signal generated which exceeds the false threshold value will be displayed as a "FALSE
CONTAM" (false contamination) and does not count towards a successful test.
Press the [ENTER] key, and then the [ ] and [ ] to adjust the false threshold value.
Scope:
0 to 99990
Default:
Twice the default working threshold
Notes:
This entry only appears if PV Test is enabled in the OPTIONS menu.
The manual adjustment increment or decrement is in steps of 10 units.
The value set cannot be lower than the generated working threshold.
20

FALSE THRESHOLD

*FLS THRSH
SETUP
CALIBRATE
I S C
2000
ENTER
RESULTS
RUN DETECT
CALIBRATE
SYSTEM FAULT
RUN ERROR