Section 7: Sweep Operation
NOTE
6. From the CONFIGURE SWEEPS menu, select SWEEP-COUNT, press the ENTER key, and
then select FINITE or INFINITE.
7. From the CONFIGURE SWEEPS menu, select SOURCE-RANGING, press the ENTER key,
and then select BEST-FIXED, AUTO-RANGE, or FIXED.
8. Press the EXIT key to return to the normal display.
Step 3. Set the delay.
Set the source delay as follows:
1. Press the CONFIG key, and then depending on the selected source channel, press the SRC1
or SRC2 key.
2. Select DELAY, and then press the ENTER key.
3. Set the delay value, and then press the ENTER key.
4. Press the EXIT key to return to the normal display.
Step 4. Turn the output on.
Press the ON/OFF OUTPUT key to turn the output on (blue OUTPUT indicator turns on). The
Model 6482 will output the programmed bias level.
Step 5. Run the sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off by
pressing the ON/OFF OUTPUT key.
7-10
The Model 6482 limits the number of sweeps (sweep-count) according to the number of
steps required for each sweep. The maximum number of finite sweeps that can be
performed is determined as follows: Maximum finite sweep count = 3000 / # points in
sweep.