Keithley 6482 Reference Manual page 208

Dual-channel picoammeter / voltage source
Table of Contents

Advertisement

Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Table 16-10
TRIGger command summary
Command
:INITiate[:IMMediate]
:ABORt
:ARM
[:SEQuence[1]]
[:LAYer[1]]
:COUNt <n>
:COUNt?
:SOURce <name>
:SOURce?
:TIMer <n>
:TIMer?
[:TCONfigure]
:DIRection <name>
:DIRection?
[:ASYNchronous]
:ILINe <n>
:ILINe?
:OLINe <n>
:OLINe?
:OUTPut <name>
:OUTPut?
:TRIGger
[:SEQuence[1]]
:COUNt <n>
:COUNt?
:DELay <n>
:DELay?
:SOURce <name>
:SOURce?
6482-901-01 Rev. A / August 2012
Description
Initiate source and measure cycle(s).
Reset trigger system. Goes to idle state.
Path to program Arm Layer:
Specify arm count (1 to 3000 or INFinite).
Query arm count (INFinite = +9.9e37).
Specify control source (IMMediate, TIMer,
MANual, BUS, TLINk, NSTest, PSTest,
BSTest).
Query control source.
Set timer interval in seconds (0.001 to 99999.99).
Query timer interval.
Enable (SOURce) or disable (ACCeptor)
bypass.
Query state of bypass.
Configure output triggers:
Select input trigger line (1, 2, 3, 4, 5, or 6).
Query input trigger line.
Select output trigger line (1, 2, 3, 4, 5, or 6).
Query output trigger line.
Output trigger (TRIGger or NONE).
Query arm output trigger status.
Path to program Trigger Layer:
Specify trigger count (1 to 3000).
Query trigger count.
Specify trigger delay: 0 to 999.9999 (sec).
Query source delay.
Specify control source (IMMediate or TLINk).
Query control source.
Return to
Section Topics
Section 16: SCPI Command Reference
Default
parameter SCPI
1
IMMediate
0.1
ACCeptor
1
2
NONE
1
0
IMMediate
16-15