Section 17: Calibration
Figure 17-10
Channel 1 current calibration connections (20 μA to 20 mA)
Model 7078-TRX-BNC
3-slot male triax to
BNC adapter
Channel 1
INPUT
CHANNEL 1
Step 6. Channel 2 current calibration adjustment
Follow the steps below to calibrate all current ranges of channel 2.
calibration ranges and currents.
1. Connect the Model 6430 to the Model 6482 INPUT CHANNEL 2 connector, as shown in
Figure
2. Select the Model 6430 V/R amperes function, set the Model 6430 output to +1.90000 nA, and
then turn on the output.
3. Press the Model 6482 MSR2 key, and then set channel 1 to the 2 nA current range using the
RANGE
4. Press the MENU key, and then select CAL and press the ENTER key.
5. Select EXECUTE, and then press the ENTER key to enter the CAL EXECUTION menu.
6. Select I-CAL and press the ENTER key. The instrument will display the following:
I-CAL CHANNEL SELECT
CHAN-1
7. Select CHAN-2 and press the ENTER key. The instrument will prompt for the positive full-range
current:
I (+FS)
Use
8. Set the Model 6430 current to +1.90000 nA.
9. Use the EDIT keys to adjust the Model 6482 display value to match the Model 6430 current.
17-26
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Model 4801
low noise
HI
coaxial cable
LO
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
INPUT
MADE IN
RATINGS MAX.
U.S.A.
30 V @ 20mA
OUTPUT ENABLE
RATINGS MAX.
COMMON
MODE
200V
CHANNEL 2
RS-232
!
TRIGGER LINK
CAT
I
LINE RATING
50, 60Hz
ANALOG
60 VA MAX
OUT
RATINGS MAX.
FUSE
LINE
30 V @ 20mA
630 mAT
100 VAC
(SB)
120 VAC
!
VOLTAGE
315 mAT
220 VAC
SOURCE
(SB)
240 VAC
OUT
30 V @ 20mA
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 6482
17-11.
key.
CHAN-2
2
= +2.000000 nA
,
,
,
, ENTER or EXIT.
Return to
Model BG-18
Dual banana to
BNC coaxial
adapter. Connect
output LO to shield.
Output
VΩA
Fluke Model 5700A multifunction calibrator
SERIAL NO.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
Section Topics
Table 17-9
summarizes
6482-901-01 Rev. A / August 2012