Figure 12-2
IEEE-488 connections
Model 707B/708B
To avoid possible mechanical damage, stack no more than three connectors on any one
instrument.
NOTE
To connect the Model 6482 to the IEEE-488 bus, follow these steps:
1. Align the cable connector with the connector located on the rear panel. The connector is
designed so that it will fit only one way.
tor.
2. Tighten the screws securely, making sure not to overtighten them.
Figure 12-3
IEEE-488 and RS-232 connector location
3. Connect any additional connectors from other instruments as required for your application.
4. Make sure the other end of the cable is properly connected to the controller. Most controllers
are equipped with an IEEE-488 style connector, but a few may require a different type of con-
12-4
Instrument
To minimize interference caused by electromagnetic radiation, use only shielded IEEE-
488 cables. Available shielded cables from Keithley are Models 7007-1 and 7007-2.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
INPUT
MADE IN
RATINGS MAX.
U.S.A.
30 V @ 20mA
RATINGS MAX.
COMMON
MODE
200V
CHANNEL 1
CHANNEL 2
!
TRIGGER LINK
CAT
I
ANALOG
OUT
RATINGS MAX.
30 V @ 20mA
VOLTAGE
SOURCE
OUT
30 V @ 20mA
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Return to
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Model 6482
Figure 12-3
shows the location of the IEEE-488 connec-
IEEE-488
RS-232
connector
connector
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
OUTPUT ENABLE
RS-232
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
100 VAC
(SB)
120 VAC
!
315 mAT
220 VAC
(SB)
240 VAC
Model 6482
Section Topics
Controller
SERIAL NO.
IEEE-488
6482-901-01 Rev. A / August 2012