Section 17: Calibration
Table 17-12
Channel 1 remote voltage bias calibration summary
Voltage bias
range
Step 4. Channel 2 input offset voltage calibration adjustment
1. Install a triaxial shielding cap on the channel 2 INPUT connector.
2. Send this command to calibrate channel 2 input offset voltage:
:CAL:PROT:SOUR2:OFFS
Step 5. Channel 2 voltage source calibration adjustment
Follow these steps to calibrate the channel 2 voltage bias source.
channel 1 calibration voltages and commands.
1. Connect the DMM to the Model 6482 channel 2 OUTPUT and INPUT LO terminals, as shown
in
Figure
2. Select the DMM DC voltage function, and enable autorange.
3. Send this command to turn on the channel 2 output:
:OUTP2 ON
4. Send the following command to select the channel 2, 10 V voltage bias range:
:SOUR2:VOLT:RANG 10
5. Send the following command to output +10 V:
:SOUR2:VOLT 10
6. Note and record the DMM reading, and send that value as the parameter for the following
command:
:CAL:PROT:SOUR2 <DMM_Reading>
17-34
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Calibration
voltages
Calibration commands
:OUTP1 ON
10 V
:SOUR1:VOLT:RANG 10
+10 V
:SOUR1:VOLT 10
:CAL:PROT:SOUR1 <DMM_Reading>
0 V
:SOUR1:VOLT 0
:CAL:PROT:SOUR1 <DMM_Reading>
-10 V
:SOUR1:VOLT -10
:CAL:PROT:SOUR1 <DMM_Reading>
30 V
:SOUR1:VOLT:RANG 30
+30 V
:SOUR1:VOLT 30
:CAL:PROT:SOUR1 <DMM_Reading>
0 V
:SOUR1:VOLT 0
:CAL:PROT:SOUR1 <DMM_Reading>
-30 V
:SOUR1:VOLT -30
:CAL:PROT:SOUR1 <DMM_Reading>
:OUTP1 OFF
17-8.
Return to
Section Topics
Table 17-13
summarizes
6482-901-01 Rev. A / August 2012