Section 17: Calibration
Step 6. Channel 1 current calibration adjustment
Follow these steps to calibrate channel 1 current measurements.
channel calibration currents and commands.
Table 17-14
Calibration currents and commands
Connections
Channel 1:
Figure 17-9
Channel 2:
Figure 17-11
Channel 1:
Figure 17-10
Channel 2:
Figure 17-12
1. Connect the Model 6430 to the Model 6482 channel 1 INPUT connector, as shown in
Figure
2. Select the Model 6430 V/R amperes function and set the output current to 1.90000 nA, and
then turn on the Model 6430 output.
3. Send the following command to select the channel 1, 2 nA range:
:SENS1:CURR:RANG 2e-9
4. Send the following command to calibrate the positive full-range current point:
:CAL:PROT:SENS1 1.9e-9
5. Set the Model 6430 current output to 0 nA.
6. Send the following command to calibrate the zero current point:
:CAL:PROT:SENS1 0
7. Set the Model 6430 current to -1.90000 nA.
17-36
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Current range
2 nA
20 nA
200 nA
2 μA
20 μA
200 μA
2 mA
20 mA
17-9.
Return to
Model 6430
Model 5700A
calibration
calibration
currents
currents
+1.90000 nA
0.00000 nA
-1.90000 nA
+19.0000 nA
0.0000 nA
-19.0000 nA
+190.000 nA
0.000 nA
-190.000 nA
+1.90000μA
0.00000μA
-1.90000μA
-
+19.0000μA
-
-
-19.0000μA
+190.000 μA
-
-
-190.000 μA
-
-
+1.90000 mA
-
-
-1.90000 mA
-
+19.0000 mA
-
-
-19.0000 mA
Section Topics
Table 17-14
summarizes
Calibration commands
:SENS1:CURR:RANG 2e-9
-
:CAL:PROT:SENS1 1.9e-9
-
:CAL:PROT:SENS1 0
-
:CAL:PROT:SENS1 -1.9e-9
:SENS1:CURR:RANG 2e-8
-
:CAL:PROT:SENS1 1.9e-8
-
:CAL:PROT:SENS1 0
-
:CAL:PROT:SENS1 -1.9e-8
:SENS1:CURR:RANG 2e-7
-
:CAL:PROT:SENS1 1,9e-7
-
:CAL:PROT:SENS1 0
-
:CAL:PROT:SENS1 -1.9e-7
:SENS1:CURR:RANG 2e-6
-
:CAL:PROT:SENS1 1.9e-6
-
:CAL:PROT:SENS1 0
-
:CAL:PROT:SENS1 -1.9e-6
:SENS1:CURR:RANG 2e-5
:CAL:PROT:SENS1 1.9e-5
0.0000μA
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-5
:SENS1:CURR:RANG 2e-4
:CAL:PROT:SENS1 1.9e-4
0.000 μA
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-4
:SENS1:CURR:RANG 2e-3
:CAL:PROT:SENS1 1.9e-3
0.0000 mA
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-3
:SENS1:CURR:RANG 2e-2
:CAL:PROT:SENS1 1.9e-2
0.0000 mA
:CAL:PROT:SENS1 0
:CAL:PROT:SENS1 -1.9e-2