Section 13: Status Structure
Measurement event register
The used bits of the measurement event register
•
Bit B0, Limit 1 Summary (L1): Set bit indicates Limit 1 summary information is available in
the :FORM:ELEM STAT element (see
•
Bit B1, Limit 2 Summary (L2): Set bit indicates Limit 2 summary information is available in
the :FORM:ELEM STAT element.
•
Bit B2, Limit 3 Summary (L3): Set bit indicates Limit 3 summary information is available in
the :FORM:ELEM STAT element.
•
Bit B3, Limit 4 Summary (L4): Set bit indicates Limit 4 summary information is available in
the :FORM:ELEM STAT element.
•
Bit B4, Limit Fail High (LFH): Set bit indicates that one of the high limit tests has failed.
•
Bit B5, Limits Pass (LP): Set bit indicates that all limit tests passed.
•
Bit B6, Reading Available (RAV): Set bit indicates that a reading was taken and
processed.
•
Bit B7, Reading Overflow (ROF): Set bit indicates that the current reading exceeds the
selected measurement range of the Model 6482.
•
Bit B8, Buffer Available (BAV): Set bit indicates that there are at least two readings in the
buffer.
•
Bit B9, Buffer Full (BFL): Set bit indicates that the trace buffer is full.
•
Bit B10: Not used.
•
Bit B11, Output Enable Asserted (OE): Set bit indicates that the output enable line is at
digital low (asserted). Source outputs can be turned on.
•
Bit B12: Not used.
•
Bit B13, Source 1 Compliance (S1C): Set bit indicates that the channel 1 voltage source is
in compliance.
•
Bit B14, Source 2 Compliance (S2C): Set bit indicates that the channel 2 voltage source is
in compliance.
•
Bit B15: Not used.
13-12
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section
Return to
Section Topics
(Figure
13-6) are described as follows:
16,
"FORMat
subsystem").
6482-901-01 Rev. A / August 2012