Keithley 6482 Reference Manual page 176

Dual-channel picoammeter / voltage source
Table of Contents

Advertisement

Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Figure 13-6
Measurement event status
STAT:MEAS:COND?
STAT:MEAS?
OR
To Measurement
Summary Bit
(OSB) of Status
Byte Register
STAT:MEAS:ENAB <NRf>
STAT:MEAS:ENAB?
6482-901-01 Rev. A / August 2012
S2C
S1C
OE
(B15)
(B14)
(B13)
(B12)
(B11)
(B10)
S2C
S1C
OE
(B10)
(B15)
(B11)
(B14)
(B13)
(B12)
&
&
&
&
OE
S2C
S1C
(B15)
(B10)
(B14)
(B13)
(B12)
(B11)
S2C = Source 2 in compliance
S1C = Source 1 in compliance
OE = Output enable asserted
BFL = Buffer full
BAV = Buffer available
ROF = Reading overflow
RAV = Reading available
Return to
Section Topics
BFL
BAV
ROF
RAV
LP
LFH
(B9)
(B8)
(B7)
(B6)
(B5)
(B4)
BFL
BAV
ROF
RAV
LP
LFH
(B9)
(B8)
(B7)
(B6)
(B5)
(B4)
&
&
&
&
&
&
BFL
BAV
ROF
RAV
LP
LFH
(B9)
(B8)
(B7)
(B6)
(B5)
(B4)
LP = Limits pass
LFH = Limit failed high
L4 = Limit 4 summary
L3 = Limit 3 summary
L2 = Limit 2 summary
L1 = Limit 1 summary
&
= Logical AND
OR = Logical OR
Section 13: Status Structure
Measurement
L4
L3
L2
L1
Condition
(B3)
(B2)
(B1)
(B0)
Register
Measurement
L4
L3
L2
L1
Event
(B3)
(B2)
(B1)
(B0)
Register
&
&
&
&
Measurement
L4
L3
L2
L1
Event Enable
(B3)
(B2)
(B1)
(B0)
Register
13-13