Section 16: SCPI Command Reference
Table 16-5 (cont.)
SENSe command summary
Command
:SENSe2
:CURRent[:DC]
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:ULIMit <n>
:ULIMit?
:LLIMit <n>
:LLIMit?
:NPLCycles <n>
:NPLCycles?
:AVERage
:COUNt <n>
:COUNt?
[:STATe] <b>
[:STATe]?
:TCONtrol <name>
:TCONtrol?
:ADVanced
:NTOLerance
<NRf>
:NTOLerance?
[:STATe] <b>
[:STATe]?
:MEDian
:RANK <NRf>
:RANK?
[:STATe] <b>
[:STATe]?
16-8
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Description
Sense 2 subsystem to control channel 2 measurement:
Path to configure current:
Configure measurement range:
Select range by specifying the expected
current reading; 0 to ±21e-3.
Query range.
Enable or disable autorange.
Query autorange.
Set autoranging upper limit (-0.021 to 0.021).
Query autoranging upper limit
Set autoranging lower limit (-0.021 to 0.021).
Query autoranging lower limit.
Specify integration rate (in line cycles): 0.01
to 10.
Query integration rate.
Path to configure moving and repeat filters:
Specify filter count; 1 to 100.
Query filter count.
Enable or disable digital filter.
Query state of digital filter.
Select filter control (MOVing or REPeat).
Query filter control mode.
Configure the advanced filter:
Set filter noise window (in %); 0 to 105.
Query filter noise window setting.
Enable or disable advanced filter.
Query state of advanced filter.
Configure and control the median filter:
Specify median filter rank; 1 to 5.
Query median filter rank.
Enable or disable median filter.
Query state of median filter.
Return to
Section Topics
Default
parameter SCPI
2e-4
ON
2e-2
2e-9
1
10
OFF
REPeat
5
OFF
1
OFF
6482-901-01 Rev. A / August 2012