Section 4: Range, Digits, Speed, and Filters
Filter programming example
Table 4-7
Average filter: Off
Median filter state: On
Median filter rank: 5
Average filter state: On
Average filter type: Moving
Average filter count: 20
Advanced filter state: On
Advance filter tolerance: 10 percent
Table 4-7
Filter programming example
Command
:MED:RANK 5
:MED ON
:AVER:TCON MOV
:AVER:COUN 20
:AVER ON
:AVER:ADV OFF
:AVER:ADV:NTOL 10
4-12
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
summarizes the command sequence to program channel 1 filter attributes as follows:
Description
Set median rank to 5.
Enable median filter.
Select moving filter type.
Set moving count to 20.
Enable average filter.
Disable advanced filter.
Set filter tolerance to 10%.