A1 alignment
Thermo Fisher Scientific
Warning Pogo pins are easily damaged by side pressure. Be careful not
to bend them.
During shipment or movement of the Multidrop Pico dispenser, it is
possible the position of the stage assembly relative to the cassette may
check
change. After re-locating the Multidrop Pico dispenser, you can follow
this procedure to test and modify the alignment of stage assembly to
cassette. If you notice poor CVs or fluids dispensed on side walls or ribs
of the destination plates you may also want to test alignment.
Please use this section only if you are comfortable with resetting
computer parameters for instruments such as the Multidrop Pico
dispenser. If you are unsure about this procedure at any point, call the
Service Help Desk for assistance. The contact information can be
found in the Multidrop Pico dispenser
1.
A1 Alignment dialog opens as shown below. Select the well plate type
that you are aligning.
Tip It is strongly recommended to verify the 1536-well plate A1 before
dispensing into 1536-well plates
2.
verification and (optional) alignment process. To skip guided
alignment, select the Manual Adjustment tab to adjust the dispense
location.
Click A1 Alignment on the Advanced tab of the ribbon. The
Press Start Guided Alignment to be guided through the
Thermo Scientific™ Multidrop™ Pico 8 Digital Dispense Operating Manual
Preventative maintenance procedures
Service and Support
document.