Dräger Caleo Instructions For Use Manual page 86

Neonatal incubator
Hide thumbs Also See for Caleo:

Advertisement

Operation
Using O2 Control
The measured value and current setpoint for O
represented by bar graphs and numerical values.
Soft key assignments:
1
= New settings are not activated.
2
off
= Switch off O
control.
2
3
on
= Switch on O
control.
2
After the system is switched on, the oxygen sensors need
a warm-up phase of at least 3 minutes. An hourglass
symbol appears in the display.
The following message is displayed at the top of the screen:
»set value with rotary knob«.
After confirming the selected O
completion of the warm-up phase, the automatic O
becomes active.
The current measured value is displayed after completing the
warm-up phase.
Adjusting the setpoint
Standard setpoint range
Extended setpoint range
Default setpoint
4
Turn rotary knob clockwise to increase setpoint.
4
Turn rotary knob counterclockwise to decrease setpoint.
4
Press rotary knob to confirm setpoint.
86
control are
2
sensor and after
2
control
2
21 vol.% to 40 vol.%
40.1 vol.% to 75 vol.%
21 %

1 2
3

4