Moving the control unit to the opposite side
for 38 mm pole
1
Loosen clamping screw to remove control unit. (Support
control unit to prevent it from dropping once clamping screw
is loosened.)
2
Loosen clamping screw to remove holder.
Upon delivery, the control unit is attached to the 38 mm dia.
tube of 310 mm length (38/310 pole).
This tube is secured for transport with a screw attachment.
If required, the tube can be detached (recommendation: ask for
assistance from qualified technical personnel):
3
Remove both caps from the tube. Insert a screwdriver into
the holes and release the tube.
To install control unit on the adjacent side:
Slide holder over the other pole.
2
Tighten clamping screw to fix holder in position.
Set control unit to the desired working height
1
Tighten clamping screw to secure control unit to the holder.
CAUTION !
When mounting control unit to its pole, ensure that there is
sufficient space to swivel the control unit.
(see "Tilting the bed", page 48).
Mounting the control unit on the handle side
CAUTION !
Only qualified technical personnel may move the control unit
from the wall side to handle side or vice versa.
Install basic pole (2M 50 680), see page 35.
Screw in the extension pole 38/600 (2M 50 691) or
pole 38/310 (2M 50 688).
Re-install the control unit as specified in the Assembly
Instructions.
CAUTION !
Do not remove cable to the control unit from the cable guides
on the basic pole.
Make sure there is sufficient space to swivel and tilt the unit.
1
2
Preparation
Installing Accessories
3
33