Dräger Caleo Instructions For Use Manual page 131

Neonatal incubator
Hide thumbs Also See for Caleo:

Advertisement

Check that access ports latch securely
Perform this test on all 4 access ports
1
Press down the ribbed area of the lever to open access port.
Close access port until lever engages in the locked position.
Try to open the access port by pulling it outwards by its edge
– it must not open.
If access port does not remain securely closed:
Take unit out of service.
Check that large access doors latches securely
Perform this test on both doors
Open door slightly.
Then, push door back into its closed position.
Turn the two knobs outwards until they engage in the
horizontal position.
WARNING !
Always ensure that both knobs of the large access doors
are engaged in position in order to avoid any risk of an
infant falling out of an incubator.
The red catch behind each knob must no longer be
visible!
WARNING !
The attached double wall must be positioned parallel to
the front door otherwise the ducted flow of warm air will
be interrupted!
If the front door fails to remain engaged or if the red marking is
visible or the double wall cannot be moved:
Take unit out of service.
Before Reusing With a Patient
1
1
2
Care
2
131