WARNING !
Always use eye protection for the infant when using
phototherapy.
WARNING !
Ensure that the ventilator circuit and all other cables,
hoses and tubing are routed correctly and safely.
Danger of extubation and disconnection!
Hoses and/or cables are at risk of being trapped when
tilting the Caleo, adjusting the height, and when opening
and closing the front door.
WARNING !
Cleaning mode must only be used when Caleo is empty.
After use, allow Caleo to cool down before dismantling.
Risk of burns when touching the heater!
WARNING !
Caleo is to be used only in rooms with line power
installations that comply with national safety standards
for hospital patient rooms (e.g., IEC/EN 60601-1 "Medical
Electrical Equipment - General Requirements For
Safety").
To maintain grounding integrity, connect only to a
"hospital grade" receptacle.
Always disconnect supply before servicing.
WARNING !
Do not use a power outlet strip for supplying power to
the Caleo incubator!
Connecting the incubator via a power outlet strip may, in
case of failure of the protective earth conductor, cause
patient leakage currents to rise above permitted limits
with a risk of electric shock to the patient.
WARNING!
Cleaning mode may only be used while Caleo is not
occupied by a patient.
WARNING!
Risk of burns upon contact with the heater!
Do not disassemble Caleo while in cleaning mode.
Important Safety Information
WARNING!
Always adjust audible alarm volume to a level that
ensures the operator will be alerted when alarms occur.
Failure to identify and correct alarm situations may
result in patient injury.
WARNING !
While alarm suppression is active, the operator of the
incubator must still assume responsibility for proper
patient care and safety in the event of an alarm.
Failure to identify and correct alarm situations may
result in patient injury.
WARNING !
Only use auxiliary electromedical equipment which
complies with national safety standards for hospital
patient rooms (e.g., IEC/EN 60601-1 "Medical Electrical
Equipment - General Requirements For Safety", UL 544).
When using the integrated power strip to connect
auxiliary devices, always observe total leakage current
and current consumption limits!
(See "Technical Data", page 166.)
WARNING !
The output of the integrated power strip is not
monitored!
Do not connect life support devices which do not have
their own power failure alarm.
WARNING !
Regularly measure infant temperature!
Do not leave canopy open for any length of time, other-
wise the air temperature inside the incubator will drop.
WARNING !
When the extended setpoint range for air temperature is
used, the infant's temperature must be continuously
monitored.
WARNING !
Monitor infant constantly when canopy, doors, or access
ports are open, to ensure infant cannot fall out of the
incubator.