Suppressing Alarms
1
When pressing this key while an alarm is active:
The audible alarm is silenced.
The central alarm indicator goes out.*
1
The LED in the button is not lit.
The maximum duration for suppressing an alarm depends on
the type of the alarm. Suppression of an alarm ends with the
alarm, and does not continue on as preemptive silencing of
alarms.
1
Pressing this key if no alarm is active (preemptive alarm
silencing):
For a period of 4 minutes, if alarms occur due to deficient
values of the following parameters:
Air temperature,
Skin temperature,
Humidity,
O
concentration,
2
no audible alarm will sound and the central alarm indicator light
will not light up*.
However: The respective alarm messages will be displayed on
screen, the measured value will start flashing, and the bar LED
will start blinking.
1
The LED in the button will be lit.
WARNING !
While alarm suppression is active, the operator of the
incubator must still assume responsibility for proper
patient care and safety in the event of an alarm.
Failure to identify and correct alarm situations may
result in patient injury.
NOTE: Alarms that had already been silenced previously, will,
regardless of their remaining silencing period, be muted also
for the duration of 4 minutes.
*
The central alarm indicator light may be switched off.
See "Configuring system settings", page 106.