Vision™ pump door and remove the administration set from the sterile packaging
leaving the end caps on the line.
2.
Connect the administration set to the medication bag.
3.
Hold the section of tubing with the set key (small plastic block)..
4.
Insert the IV tubing into the pump by placing the set key into the keyway as shown
by the arrow in the diagram. Insert the tubing from left to right, and avoid stretching
or pulling the tubing. Check that the key located on the tube is placed in its correct
position in the tubing guide. The anti-stretch locator (blue ring) is placed in the
round space to the left of the set key. The flow direction arrows inside the pump
door indicate flow direction from the bag to the patient.
NOTE: the set key can only be fitted into the keyway one way. If you are having
troubles fitting the key into the keyway, do not force it. Turn the line around to
ensure you have correctly lined up with the direction of flow.
5.
For products equipped with Anti-Stretch locator, locate the round blue Anti-Stretch
locator in the round space on the left side.
6.
Verify that the distal flow valve is on the right hand side of the pump.
7.
Close the pump door until the latch clicks.
NOTE: ensure that the tubing is inserted completely into the pumping channel.
Warning: if the sets sterilized package is damaged, it should not be used.
Warning: use aseptic technique. Patient infection may result from the use of
non-sterile components. Maintain sterility of all disposable components and use
only once single use consumables marked

Priming the Set

1.
Once the set is loaded turn on the BodyGuard 595
Color Vision™ pump, by pressing
The screen than will prompt for level one code. Enter code 700 and press
2.
BodyGuard 595 Color Vision™ Directions For Use
.