Loading The Administration Set; Priming The Line - CME VET-PRO VIP 2000 Operator's Manual

Veterinary infusion pump system
Hide thumbs Also See for VET-PRO VIP 2000:
Table of Contents

Advertisement

Loading the Administration Set

NOTE:
Follow the instructions supplied with the individual administration set.
Warning:
the accuracy of the Infusion.
1.
Lift the door latch (on the right side of the door) to open the VET-PRO VIP 2000™ pump door
and remove the administration set from the sterile packaging leaving the end caps on the line
until the moment of connection.
2.
Connect the administration set to the medication bag.
3.
Hold the section of tubing that you intend to load and make sure the flow direction is in line with
the flow direction arrows inside the pump door.
4.
Insert the IV tubing into the canal designed for it as shown by the arrow in diagram 1. Insert the
tubing from left to right with the shorter end connected to the medication bag on the left, and
avoid stretching or pulling the tubing.
5.
Close the pump door until the catch clicks.
NOTE:
Ensure that the tubing is inserted completely into the pumping canal.
Warning:
components. Maintain sterility of all disposable components and use only once single use
consumables marked

Priming the Line

Prime function is used:
1.
Before starting a program
2.
After air in line alarm
Warning:
NOTE:
The default priming flow rate is 600 ml/hr. During priming, the air in line alarm is
disabled.
VET-PRO VIP 2000™ Color Vision Operator Manual
Use of low quality administration set may impair the operation of the pump and
Use aseptic technique. Patient infection may result from the use of non-sterile
.
Always disconnect IV tubing from patient before starting priming procedure.
4 – Operation
27

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents