And/Or Ethernet Mode Troubleshooting Tips; Table 3-11; Table 3-12 - Xantrex ENET-XFR Operating Manual

Internal ethernet and rs-232 interface for xfr series programmable dc power supplies
Table of Contents

Advertisement

RS-232
Mode Troubleshooting Tips

Table 3-11

RS-232
S y m p t o m
Cannot establish
communication
RS-232
and/or

Table 3-12

RS-232
S y m p t o m
Instrument not
responding to new
user settings
The output is
displaying zero
volts and zero amps,
but the setpoints are
non-zero
TM-XR9B-01XN
Mode Tips
P o s s i b l e C a u s e s
Using straight through
RS-232
cable instead of
Null Modem cable
Loose cables
Baud rate mismatch
Trying to communicate via
the wrong COM port
The interface card is
configured for
Ethernet
communications
Ethernet
Mode Troubleshooting Tips
and/or
Ethernet
Mode Tips
P o s s i b l e C a u s e s
The unit is in local mode
The output is off
The output has tripped due
to an OVP, or foldback
protection event
C o r r e c t i v e A c t i o n s
Use a Null Modem cable
Ensure that the cables are securely connected to
the computer and the power supply.
Select a baud rate that matches the instruments
settings. See
"Baud Rate Selection" on page 2–
18
for details on changing this setting.
Most computers have two COM ports, COM1
and COM2, so ensure you have selected the
correct COM port.
Reconfigure the interface card for RS-232
communications. See
or RS-232" on page 2–12
C o r r e c t i v e A c t i o n s
Put the unit into remote mode by sending the
command REN ON.
Send the command OUT ON
Correct the condition that has caused the fault,
and send the command RST
Troubleshooting
"Configuring for Ethernet
for details.
3–23