Command Reference; Table 3-7 Command Reference - Xantrex ENET-XFR Operating Manual

Internal ethernet and rs-232 interface for xfr series programmable dc power supplies
Table of Contents

Advertisement

Command Reference

Table 3-7 Command Reference

C o m m a n d
D e s c r i p t i o n
ASTS?
Asks for the supply's accumulated status register. The accumulated status register
stores any bit that was entered in the status register since the accumulated status
query command (ASTS?) was last used, regardless of whether the condition still
exists. The accumulated status register has the same bits, weights, and conditions
as the status register. A bit in the accumulated status register will be set at 1 if the
corresponding bit in the status register has been 1 (TRUE) at any time since the
register was last read. See
page
Response: ASTS <status mask> where status mask is the decimal equivalent of the
total bit weights for the operating conditions as listed in the status register.
AUXA <1/ON>,
Controls the AUXA output signal level at rear panel connector Pin 7. Active low.
<0/OFF>
Initial value: AUXA 0
AUXA?
Asks for the present set value of the AUXA output signal.
Response: AUXA 0 (OFF) or AUXA 1 (ON)
AUXB <1/ON>,
Controls the AUXB output signal level at rear panel connector Pin 8. Active low.
<0/OFF>
Initial value: AUXB 0
AUXB?
Asks for the present set value of the AUXB output signal.
Response: AUXB 0 (OFF) or AUXB 1 (ON)
CLR
Initializes the power supply to its power ON condition. If issued while in local
mode, CLR will force power supply settings to register default values as in but
these default settings will not come into effect until the power supply is switched to
remote mode operation. The CLR commands will clear faults from the fault
register. CLR will not reset CMODE.
CMODE <1/ON>,
CMODE ON places the power supply into calibration mode for processing
<0/OFF>
calibration commands.
Initial value: CMODE OFF or CMODE 0
CMODE?
Asks for the power supply's calibration mode status.
Response: CMODE 0 (disabled)
CMODE 1 (enabled)
TM-XR9B-01XN
"Accumulated Status, Status, and Fault Registers" on
3–17. The ASTS? query clears the status register.
Command Reference
3–9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Enet-xfr3

Table of Contents