Current Readback Calibration Procedure - Xantrex ENET-XFR Operating Manual

Internal ethernet and rs-232 interface for xfr series programmable dc power supplies
Table of Contents

Advertisement

Calibration

Current Readback Calibration Procedure

4–6
7. Program the supply at various levels using the ISET command to
confirm that the calibration was successful and that linearity is
observed. Refer to the current program accuracy specification in
Appendix
A.
8. Turn off calibration mode by sending the command CMODE OFF or
CMODE 0 to the power supply.
1. Connect the current shunt and voltmeter to the power supply as
shown in
Figure
2. Activate calibration mode by sending command CMODE ON or
CMODE 1 to the power supply.
3. Send command IRLO; VRLO to the power supply. Wait for the
supply to settle. Measure and record the output voltage shown on the
external voltmeter. Send IRLO again.
4. Send command IRHI; VRHI to the supply. Wait for the supply to
settle. Measure and record the output voltage shown on the external
voltmeter. Send IRHI again.
5. Calculate IRLO and IRHI from the voltages taken from the external
voltmeter and the shunt resistance. I=V/R.
6. Send the command IRDAT <ilo>,<ihi> to the power supply. <ilo>
and <ihi> are the current values obtained from sending the IRLO and
IRHI commands to the power supply. When the power supply is
calibrated, the low to high current readback calibration values are
stored as constants.
7. Program the supply at various levels using the ISET command to
confirm that the calibration was successful and that linearity is
observed. Refer to the current readback accuracy specification in
Appendix
A.
8. Turn off calibration mode by sending the command CMODE OFF or
CMODE 0 to the power supply.