Calibration
4–2
You can calibrate the
Voltage program
Voltage readback
Current program
Current readback
Overvoltage protection
The following equipment will be required to accurately calibrate your
unit:
Digital Voltmeter, 5 1/2 digit, 0.1% accuracy or better, with test leads
Current sensing shunt resistor, rated for 150% of maximum output
current, 0.25% accuracy or better
Connection wires rated for the unit's maximum output current and
voltage
A network ready, or RS-232 equipped computer to send the
calibration commands to the unit
Calibrate the unit according to the following procedures, referring to
"Command Reference"
commands used.
The calibration procedures in this section are designed to be performed at
an ambient temperature of 25°C ± 5°C.
Ethernet/RS-232 Interface
for more information about the calibration
for: