Table of Contents

Advertisement

Quick Links

User's Manual
Puritan Bennett
TM
520 Ventilator

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Puritan Bennett 520 Ventilator and is the answer not in the manual?

Questions and answers

Summary of Contents for Covidien Puritan Bennett 520 Ventilator

  • Page 1 User’s Manual Puritan Bennett 520 Ventilator  ...
  • Page 2 Covidien’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided. Nothing in this manual shall limit or restrict in any way Covidien’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice.
  • Page 3: Table Of Contents

    Contents Preface ................Purpose of This Manual .
  • Page 4 Contents Connecting to an External DC Power Source........4–4 Patient Circuit .
  • Page 5 Contents Range, Resolution, and Accuracy ..........A–4 Environmental .
  • Page 6 Contents This page intentionally blank User’s Manual...
  • Page 7 Figures Figure 1-1. Locations of Labels – Top-Front View....... . . 1–16 Figure 1-2.
  • Page 8 Figures Figure 5-4. Selecting the USB Menu ..........5–4 Figure 5-5.
  • Page 9 Tables Table 1-1. Ventilator Symbols ..........1–11 Table 1-2.
  • Page 10 This page intentionally blank viii User’s Manual...
  • Page 11: Preface

    Information regarding your product warranty is available from your sales representative or Covidien. Extended Service The Puritan Bennett™ 520 Ventilator offers extended service contracts/warranties for purchase when the ventilator is purchased. Please contact your local Covidien Sales or Service Representative for additional information. For online technical support, visit the SolvIT Center Knowledge Base by clicking the link at http://www.respiratorysolutions.covidien.com.
  • Page 12: Technical Support

    Preface Technical Support Technical Service Contacts: Covidien Argentina Covidien Australia Covidien Austria GmbH Aguero 351 52A Huntingwood Drive Campus21 Capital Federal - 1171 ABC, Huntingwood, NSW 2148 Europaring F09402 Argentina Australia Brunn am Gebrige Tel: (5411) 4863-5300 Telephone (+61) 1800 350702 A-2345 Österreich...
  • Page 13 Fax + 46 858 56 05 29 Tel +41 17865050 Fax +34 93 373 87 10 Fax +41 17865010 Covidien UK and Ireland Asia: Tyco Healthcare Pte Ltd Unit 2, Talisman Business Park Singapore Regional Service Centre London Road, Bicester...
  • Page 14 This page intentionally blank Preface-4 User’s Manual...
  • Page 15: Safety Information

    Definitions 1 Safety Information Definitions This manual uses three indicators to highlight critical information: Warning, Caution, and Note. They are defined as follows: WARNING Indicates a condition that can endanger the patient or the ventilator operator. Caution Indicates a condition that can damage the equipment. Note: Indicates points of particular emphasis, that make operation of the ventilator more efficient or convenient.
  • Page 16 If the ventilator fails the alarm tests or if you cannot complete the tests, refer to section 3, “Alarms and Troubleshooting” or call your equipment supplier or Covidien. • The default setting for altitude compensation is YES. Altitude compensation should always be set to YES for accurate volume delivery calculations at all elevations.
  • Page 17 The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored to verify normal...
  • Page 18 Safety Information • When changing the mode during ventilation, significant transitions of pressure, flow or cycling rate might occur, depending on the difference between the modes. Before setting the new mode, first ensure that the settings between the different modes are compatible. This reduces the risk of discomfort and harm to the patient.
  • Page 19 Warnings • When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator. • Adjustable alarms should not be systematically cancelled; instead, they should be adjusted according to the needs and condition of the patient.
  • Page 20: Safety Information

    Safety Information Warnings regarding Electrical Power • To ensure that ventilation continues uninterrupted, ensure alternative power sources are available (AC power source, extra batteries, or an auxiliary DC car adapter). Be prepared for the possibility of power failure by having an alternative means of ventilation ready for use. •...
  • Page 21 • If using oxygen with the Puritan Bennett™ 520 Ventilator, Covidien recommends using an oxygen analyser with minimum and maximum concentration alarms. Always measure the delivered oxygen with a calibrated oxygen analyser that features a minimum and maximum concentration alarm in order to ensure that the prescribed oxygen concentration is delivered to the patient.
  • Page 22 Safety Information • To ensure stability, when the Puritan Bennett™ 520 Ventilator is mounted on a cart, the weight of the oxygen bottle should not exceed 14 kg (30 lbs). • The oxygen supply hose ages even when it is not in use and should be replaced periodically. The expiration date may be located on the back of the hose end-piece.
  • Page 23: Humidifier

    The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored to verify normal...
  • Page 24 • To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only personnel authorised and qualified by Covidien should attempt to service or make authorised modifications to the Puritan Bennett™ 520 Ventilator.
  • Page 25: Symbols And Markings

    Symbols and Markings Warnings to Protect the Environment • Batteries should be disposed of according to environmental legislation in your country and locality. • For environmental protection, the ventilator and its components, whatever their respective conditions of operation, cannot be disposed of with household waste and must be submitted for suitable selective collection and possible recycling of the materials.
  • Page 26 Safety Information Table 1-1. Ventilator Symbols Symbols Descriptions This symbol appears on the ventilator’s front panel UP key; see Figure 2-3 on page 2-6, item 4. This key is used to: move the LCD display’s cursor upwards, line-by-line; increase the value of displayed and selected parameter setting.
  • Page 27 Symbols and Markings Table 1-1. Ventilator Symbols Symbols Descriptions Internal Battery. This symbol appears on the top-center of ventilator’s LCD display to indicate that the ventilator is being powered by its internal battery. See Figure 2-4 on page 2-7, item 1 and refer to chapter 6, “Internal Battery”, for more information.
  • Page 28 This symbol indicates a communications port for interfacing with a USB connector. See Figure 1-2, item 9. PC connector. This symbol indicates a port that can be used by authorised Puritan Bennett product service personnel or Covidien service personnel for software maintenance. See Figure 1-2, item 8. Atmospheric pressure limitation. Humidity limitations.
  • Page 29: Labels / Identification And Instruction Information

    Labels / Identification and Instruction Information Labels / Identification and Instruction Information Various labels or specific markings are affixed to the ventilator that describe precautions to be taken for the cor- rect use of the ventilator and contribute to the traceability of the product. Refer to Table 1-2 and the figures on the following pages for illustrations of these labels and markings and their locations on the ventilator.
  • Page 30: Figure 1-1. Locations Of Labels - Top-Front View

    Safety Information Note: The item number callouts in the following figures refer to those listed in Table 1-2. Figure 1-1. Locations of Labels – Top-Front View Figure 1-2. Location of Labels and Markings – Rear View 1-16 User’s Manual...
  • Page 31: Figure 1-3. Location Of Labels - Bottom View

    Labels / Identification and Instruction Information Figure 1-3. Location of Labels – Bottom View 1-17 User’s Manual...
  • Page 32 Safety Information This page intentionally blank 1-18 User’s Manual...
  • Page 33: Ventilator Overview

    Indications for Use 2 Ventilator Overview Indications for Use The Puritan Bennett™ 520 Ventilator is indicated for the continuous or intermittent mechanical ventilatory support of patients weighing at least 11 lb (5 kg) who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a doctor.
  • Page 34: Contraindications

    Ventilator Overview • Homecare providers • Patient and patient’s families WARNING This ventilator must be used only under the responsibility and on the prescription of a doctor. The ventilator is not intended to be used for patients without breathing autonomy or who are ventilator dependent.
  • Page 35: Breathing Circuit

    Device Classification WARNING To reduce the risk of a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (such as flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses. Oxygen therapy for patients with respiratory failure is a common and effective medical prescription. However, be aware that inappropriate oxygen use may potentially lead to serious complications, including, but not limited to, patient injury.
  • Page 36: Front Panel

    Ventilator Overview Front Panel LCD Display – Displays information about the ventilator Patient Pressure Monitoring Port – Nipple for including patient hours and software version, ventilation monitoring proximal patient pressure. modes and settings, and monitored and calculated patient data. The display also allows the user to view and, using the Control Panel, adjust the ventilator’s operating and alarm configuration settings.
  • Page 37: Back Panel

    Back Panel Back Panel Ergonomic carrying handle. PC Cable Connector: USB mini-B connector used for Puritan Bennett™ Ventilator Test Software. On/Off (I/O) switch with protective cover: Inlet Port: Device powered on in position I; device switched Connects the ventilator to a low pressure oxygen off in position 0.
  • Page 38: Control Panel

    Ventilator Overview Control Panel Alarm indicators (two LEDs): DOWN key: • Red indicator: Moves the cursor down and decreases parameter values. • Continuous: Very High Priority (VHP) alarm activated • High priority (HP) alarm activated. Yellow indicator: • Medium priority (MP) alarm activated. ALARM CONTROL key: MENU key: •...
  • Page 39: Ventilation Menu

    Ventilation Menu Ventilation Menu Ventilation menu with ventilation on standby. Ventilation menu during ventilation. General information line: Ventilation settings: Preferences menu access line: Displays the current ventilation mode, Displays the specific ventilation Highlight this line and press along with the following: parameter values for the the ENTER key to display...
  • Page 40: Alarm Menu

    Ventilator Overview Alarm Menu Alarm menu with ventilation on standby. Alarm menu when not in standby. Title line: Alarm settings: Access line to Alarm Log menu. Displays ventilation mode and the Displays the specific alarm parameter following symbols: values for the currently selected Highlight this line and ventilation mode, which are: press the ENTER...
  • Page 41: Usb Memory Device Menu

    Keep in mind that troubleshooting information is available in this manual to assist you in the event of a problem. Refer to chapter 3, “Alarms and Troubleshooting”. If you cannot determine the cause of a problem, contact your equipment supplier or Covidien. Refer to Appendix 8.3, “Service Assistance”.
  • Page 42 Ventilator Overview This page intentionally blank 2-10 User’s Manual...
  • Page 43: Alarms And Troubleshooting

    3 Alarms and Troubleshooting WARNING When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem, examine the patient first before examining the ventilator. Alarm volume should be adjusted with respect to the ventilator’s operating environment and so that the patient's caretakers can hear the alarms.
  • Page 44: Alarm Level Of Priority

    Alarms and Troubleshooting Alarm Level of Priority The alarm hierarchy for signalling the level of alarm criticality is listed below. • Very High Priority (VHP): Immediate critical situation; ventilation is impossible: Continuous Sound Signaling / With or Without Continuous Red LED Illumination / With or Without Message / With or Without Display Lighting (it is possible for an alarm condition to occur that may not have both a message and lighting).
  • Page 45: Alarm Display

    Alarm Display Alarm Display During operation, when an alarm is activated: • One of the red or yellow alarm indicators to the left of the ALARM CONTROL key illuminates and flashes. • An alarm tone sounds. • A message is displayed and flashes in reverse video at the bottom of the Ventilation Menu or Alarm Menu.
  • Page 46: Alarm Log Menu

    Alarms and Troubleshooting Alarm Log Menu All alarms are recorded in the internal memory of the ventilator at the time they are activated. The Alarm Log menu is used to display the last eight (8) alarms activated, along with their date and time of activation.
  • Page 47: Silencing The Audible Portion Of Alarms

    Silencing the Audible Portion of Alarms Figure 3-4. Alarm Log Display when No Alarm Activated To dismiss the Alarm Log screen manually: Press the ENTER key when the cursor is on the “Back” line. The Alarm Log screen is dismissed automatically: •...
  • Page 48: Pausing/Resetting Alarms

    Alarms and Troubleshooting The audible portion of activated alarms is automatically reactivated: • After 60 seconds, if the cause(s) of the alarm(s) persist(s) • Whenever a new alarm is activated Note: If a key is stuck or held down for 45 seconds a keypad alarm will occur. Pausing/Resetting Alarms WARNING When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or problem,...
  • Page 49: Re-Activating Alarms

    Re-activating Alarms • The visual alarms will be reset. Re-activating Alarms Alarms that have been paused and whose activation conditions continue to exist can be reactivated. To reactivate alarms, proceed as follows: 1. Press the MENU key to access the Alarm Setting menu, if this is not the menu currently displayed. 2.
  • Page 50: Overview Of Alarms

    Alarms and Troubleshooting Overview of Alarms Note: The message: “*IF PERSISTS RESTART/SRVC” will occur for certain alarms only if the alarm condition continues for longer than 30 seconds. These alarms are listed in See Table 3-1, Overview of Alarms and See Table 3-2, Alarms and Corrective Actions.
  • Page 51 Overview of Alarms Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Buzzer Battery Failure. The Battery Buzzer Voltage is too BUZZER FAULT4 low. RESTART/SRVC Internal technical problem that prevents the battery sounding the POWER SUPPLY LOSS alarm.
  • Page 52 Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. No exhalation valve connected with PEEP set to less than 4 mbar CONNECT VALVE OR CHANGE PRESS PIP set to more than 30 mbar when relative pressure is set to OFF.
  • Page 53 Overview of Alarms Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Internal battery capacity < 10 min. or 3%. (battery voltage < 22.5V) EMPTY BATTERY Consequence: ventilation comes to a halt. Battery temperature out of HIGH / LOW BATTERY tolerance.
  • Page 54 Alarms and Troubleshooting Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Keyboard key held down for KEYPAD FAULT more than 45 seconds. RESTART/SRVC* *IF PERSISTS RESTART/SRVC Internal battery capacity LOW BATTERY <...
  • Page 55 Overview of Alarms Table 3-1. Overview of Alarms (Continued) Audio Alarm Paused Paused Alarm Message Cause/Ventilator Response Priority Avail. Avail. Detection of a fault in the POWER FAULT electrical power supply system. RESTART/SRVC 1. Electrical power supply to the machine is interrupted with the I/O switch when ventilation is in progress 2.
  • Page 56: Troubleshooting

    To ensure proper servicing and avoid the possibility of physical injury to personnel or damage to the ventilator, only personnel authorised and qualified by Covidien should attempt to service or make authorised modifications to the Puritan Bennett™ 520 Ventilator.
  • Page 57 Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom Defective operation of the buzzers. Restart ventilator to see if alarm clears. If BUZZER FAULT1 not, replace the ventilator and call your Consequence: no audible tone when an RESTART/SRVC customer service representative.
  • Page 58 Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom No connection of the proximal pressure Reconnect the proximal pressure line. tube when ventilation starts. Reconnect the connection line or replace it if obstructed.
  • Page 59 Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom If patient has been disconnected, DEVICE FAULT9 reconnect patient to reset the fault. POST RAM Error. RAM Read/Write does not IF PERSISTS If persists restart ventilator to see if alarm match memory setting.
  • Page 60 Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom CAUTION: Ensure that ventilator is being used according to the operating instructions found in Appendix A, “Specifications”. If the ambient temperature is too low, place the device in a warmer environment.
  • Page 61 Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom Check patient’s trachea and clear the Airway obstruction. obstruction. If the filter is obstructed, replace the filter. Proximal pressure tube or patient circuit Clean the proximal pressure tube or the obstructed.
  • Page 62 Alarms and Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom OCCLUSION CHECK CIRCUIT Clean, unblock, and/or properly connect Patient circuit obstructed. the patient circuit. *IF PERSISTS RESTART/SRVC A non-vented configuration is being used or the built-in leak in the mask or in the Replace the non-vented circuit with a circuit may be obstructed or insufficient for...
  • Page 63: Additional Troubleshooting

    Troubleshooting Table 3-2. Alarms and Corrective Actions (Continued) Alarm Message or Possible Reason(s) For The Alarm Event Corrective Action(s) Symptom The ventilation settings are not compatible Remove exhalation valve to start CPAP REMOVE VALVE with the type of patient circuit used. ventilation.
  • Page 64 Alarms and Troubleshooting Table 3-3. Additional Troubleshooting and Corrective Actions (Continued) Conditions Possible Causes Corrective Actions Adjust contrast or call your customer service representative if the problem Unusual display persists. Problem with the display unit. on the screen Ensure that the ventilator is not exposed to direct radiation from the sun.
  • Page 65: Installation And Assembly

    Installing the Ventilator 4 Installation and Assembly WARNING Before operating the ventilator, read, understand, and strictly follow the information contained in Chap- ter 1, “Safety Information”. The ventilator is not intended to be used for patients without breathing autonomy or who are ventilator dependent.
  • Page 66: Connecting To External Ac Power

    The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored to verify normal operation.
  • Page 67: Figure 4-2. Inserting The Power Cable Holder Into The Notch

    Connecting to External AC Power To secure the AC power cable: 1. Insert the power cable holder (Figure 4-2, item 1) into the notch of the battery cover. Figure 4-2. Inserting the Power Cable Holder Into the Notch 2. Push the AC power cable into the power cable holder (Figure 4-2, item 2).
  • Page 68: Connecting To An External Dc Power Source

    Installation and Assembly Figure 4-4. Power Indicators To disconnect the AC power cable: 1. Disconnect the AC power cable from the AC power outlet. 2. Disconnect the AC power cable from the ventilator’s AC connector at the rear of the device. 3.
  • Page 69: Figure 4-5. Connecting The Ventilator To An External Dc Power Source

    Connecting to an External DC Power Source Then connect the DC power cable into the car auxiliary adapter. Note: Whenever AC power is unavailable, the ventilator can operate from a continuously powered external 12 – 30 VDC power source via a DC power cable (Figure 4-5, item 1) that connects to the ventilator’s rear panel DC power input connector...
  • Page 70: Patient Circuit

    After assembling, cleaning, or reassembling the patient circuit, and on a daily basis, inspect the hoses and other components to ensure that there are no cracks or leaks and that all connections are secure. To ensure proper performance of the ventilator, use a patient circuit recommended by Covidien in this manual; refer to chapter 4, “Installation and Assembly”...
  • Page 71: Installing The Patient Circuit

    Patient Circuit 4.4.2 Installing the Patient Circuit The patient circuit is mounted depending on the setup of the circuit used and the accessories used. The following procedures describe the installation of the patient circuit with a humidifier. To add other accessories, refer to the installation instructions for the specific accessories used. To connect a single limb circuit with an exhalation valve: (refer to Figure 4-7)
  • Page 72: Figure 4-7. Single Limb Patient Circuit With Exhalation Valve

    Installation and Assembly Figure 4-7. Single Limb Patient Circuit With Exhalation Valve Note: Although shown here, the humidifier (item 10), water trap (item 5), and tubes upstream of the single limb patient circuit are not included with the ventilator. Contact your supplier for more information. Figure 4-8.
  • Page 73: Figure 4-9. Single Limb Patient Circuit Without Exhalation Valve

    Patient Circuit 1. Inspect the components of the patient circuit for any signs of damage, such as cracks (which might cause leakage). Do not use damaged components to assemble the patient circuit. 2. Install the bacteria filter (item 1) on the TO PATIENT outlet port, as shown. 3.
  • Page 74: Filters

    Installation and Assembly The level of inspiratory resistance of the circuit and accessories (bacteria filter, humidifier, HMEs etc) must be as low as possible. Settings—particularly the PATIENT DISCONNECTION alarm, maximum inspired volume (Max VTI), and minimum inspired volume (Min VTI) settings—must be periodically adjusted according to changes in the patient circuit resistance—especially when filters are replaced.
  • Page 75: Bacteria Filter

    Humidifier WARNING The air inlet filter is for use on a single patient and is not reusable; do not attempt to wash, clean, or reuse Failing to replace a dirty air inlet filter, or operating the ventilator without a filter, may cause serious damage to the ventilator.
  • Page 76: Oxygen

    Installation and Assembly Figure 4-12. Humidifier When a humidification device is used, any condensation that forms in the patient circuit is collected in the water trap. If you notice any moisture in the patient circuit, you need to replace the wet circuit components with dry ones.
  • Page 77: Connecting The Oxygen Supply

    Oxygen WARNING To avoid injury to the patient and/or possible damage to the ventilator: before using the ventilator, use a flow meter (flow regulator) to regulate the oxygen supply to specifications before connecting the ventilator to the oxygen supply. Ensure that the oxygen supply pressure to the machine never exceeds 7 psi (50 kPa) or a flow of 15 lpm. Refer to Table A-8 on page A-3 for sensitivity tolerances.
  • Page 78: Figure 4-14. Connecting The Oxygen Supply System

    Installation and Assembly WARNING Before connecting the oxygen supply, ensure that the stud on the oxygen connector (Figure 4-13, item 3) is protruding outwards. Inspect the oxygen coupler (Figure 4-14, item 2) before use to ensure it has its black o-ring attached and in good condition.
  • Page 79: Fitting The Ventilator Into The Dual Bag

    Fitting the Ventilator into the Dual Bag 2. Press the locking tab of the ventilator’s oxygen connector, as shown in Figure 4-15, to unlock the oxygen connection. Figure 4-15. Disconnecting the Oxygen Supply System 3. Disconnect the oxygen supply’s oxygen connector by pulling it towards you. The ventilator’s oxygen connector’s locking stud (Figure 4-14,...
  • Page 80: Mounting The Ventilator On A Wheelchair

    Installation and Assembly Mounting the Ventilator on a Wheelchair WARNING To reduce the risk of a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (such as flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses. Due to its limited internal battery’s reserve capacity, the ventilator should only be operated on the internal battery when no other power source is available.
  • Page 81: Mounting The Ventilator On The Utility Cart

    Mounting the Ventilator on the Utility Cart Figure 4-16. Using the Dual Bag 4.10 Mounting the Ventilator on the Utility Cart Match the mounting holes (item 1) on the bottom of the Puritan Bennett™ 520 Ventilator to the mounting studs (item 2) on the top of the utility cart platform. Figure 4-17.
  • Page 82: Connecting The Nurse Call Cable

    Installation and Assembly Figure 4-18. Using the Utility Cart 4.11 Connecting the Nurse Call Cable Connect the Nurse Call cable (Figure 4-19, item 1) to the Nurse Call Monitor Connector (item 2). Figure 4-19. Connecting the Nurse Call Cable WARNING Before using the Nurse Call system, ensure that its connections are secure and it operates properly.
  • Page 83: Operating Procedures

    3.8, “Trouble- shooting” or call your equipment supplier or Covidien. Due to its limited internal battery’s reserve capacity, the ventilator should only be operated on the inter- nal battery when no other power source is available. Ensure that the internal battery never becomes fully discharged.
  • Page 84 Operating Procedures Figure 5-1. Turning on the Ventilator The following events occur: • The ventilator is powered on. • A Power On Self Test (POST) is carried out (when plugged in to an AC power source). • The front panel indicators flash (except for the indicator showing the type of power supply in use, which remains lit).
  • Page 85: Usb Menu Parameters

    The USB Menu is not accessible from the Setup Menu or Maintenance menu. To access patient data via a PC, a dedicated software package, Puritan Bennett™ Respiratory Insight Software, is available for Clinicians. Contact Covidien or your Puritan Bennett product representative for further information.
  • Page 86: Usb Menu

    Operating Procedures 5.2.2 USB Menu To access the USB menu when a USB memory device is connected: Press the MENU key several times, until the USB Menu appears: Figure 5-4. Selecting the USB Menu In case of high priority alarm activation the ventilator will automatically display the alarm page. To return to the USB Menu, press the MENU key.
  • Page 87: Transfer Trends

    USB Menu Parameters To transfer continuous data from a ventilator to a USB memory device: 1. Use the UP or DOWN arrow keys to place the cursor at the “Transfer Continuously” position. 2. Press ENTER. • The cursor changes to the plus/minus symbol. •...
  • Page 88: Starting Ventilation

    Operating Procedures To transfer trend data from a ventilator to a USB memory device: 1. Use the UP or DOWN arrow keys to place the cursor at the “Transfer Trends” position. 2. Press ENTER. • The cursor changes to the plus/minus symbol. •...
  • Page 89: Stopping Ventilation

    Stopping Ventilation When the ventilator is in standby (the ventilator is on, but ventilation has not started), a message that prompts the ventilator operator to press VENTILATION ON/OFF to start ventilation is displayed in the right-hand window of the ventilation and alarm menus (Figure 5-7).
  • Page 90 Operating Procedures To stop the ventilator: 1. Press and hold down the VENTILATION ON/OFF key (Figure 5-8, item 1) for three (3) seconds. • A message prompting the user to keep the button pressed appears on the monitoring window, as shown in the graphic below: Figure 5-9.
  • Page 91: Turning Off The Ventilator

    Turning Off the Ventilator Turning Off the Ventilator WARNING When the ventilator is switched back on, it will immediately begin ventilating—without the user first hav ing to press the VENTILATION ON/OFF key. Handle the ventilator with care after use, particularly when ambient temperatures are high. Some ventila- tor surfaces may be very hot, even if safety specifications are not exceeded.
  • Page 92 Operating Procedures This page intentionally blank 5-10 User’s Manual...
  • Page 93: Internal Battery

    Battery Capacity 6 Internal Battery WARNING Ensure that the ventilator’s internal battery is fully charged before using the ventilator with an external DC power source. Powering the ventilator using an external 12– 30 VDC power source (via the DC power cable) does not enable charging of its internal battery.
  • Page 94: Battery Operation

    Internal Battery Table 6-1. Internal Battery Reserve Capacity Displayed Values Average Operating Time on Internal Battery Power Vt 200 ml (± 5 ml) PIP  10 mbar (± 2 mbar) 5 hours (–10%) Rtot  20 bpm Vt300ml (± 5 ml) PIP ...
  • Page 95: Figure 6-2. Battery Reserve Capacity As A Percentage

    Battery Operation Figure 6-1. Internal Battery Indicator • A loss of external power supply alarm is activated. If ventilation is stopped, the internal battery reserve capacity is displayed as a percentage of battery charge. Refer to Figure 6-2. Figure 6-2. Battery Reserve Capacity as a Percentage If the ventilator is running, the internal battery reserve is momentarily displayed as a percentage.
  • Page 96: Testing The Battery

    Internal Battery WARNING Due to its limited internal battery’s reserve capacity, the ventilator should only be operated on the internal battery when no other power source is available. Ensure that the internal battery never becomes fully discharged. Alarms will sound when the charge on the internal battery is low. See Table 3-1, Overview Alarms.
  • Page 97: Storage

    Storage • The “INTERNAL BATTERY“ indicator flashes (Figure 6-4, item 2). Figure 6-4. Power Indicators When Charging the Battery When the battery charge is complete, the “INTERNAL BATTERY“ indicator turns off. WARNING In ambient temperatures above 40°C (104°F) the battery may not be fully charged even if the “INTERNAL BATTERY”...
  • Page 98 Internal Battery This page intentionally blank User’s Manual...
  • Page 99: Cleaning

    Cleaning the Ventilator 7 Cleaning WARNING A patient treated by mechanical ventilation is highly vulnerable to the risks of infection. Dirty or contaminated equipment is a potential source of infection. Clean the ventilator and its accessories regularly and systematically before and after each use and following any maintenance procedure to reduce the risks of infection.
  • Page 100: Cleaning The Accessories

    Cleaning Table 7-1. Approved Cleaning Solutions for Exterior Ventilator Surfaces (Continued) Description Hydrogen peroxide 15% ammonia (85% tap water) Ammonia-based household cleaners Household cleaners Cleaning the Accessories Follow the accessory manufacturer’s instructions for cleaning the ventilator’s accessories and components, including the patient circuit. WARNING After assembling, cleaning, or reassembling the patient circuit, and on a daily basis, inspect the hoses and other components to ensure that there are no cracks or leaks and that all connections are secure.
  • Page 101: Routine Maintenance

    Do not attempt to open, repair or otherwise service the ventilator yourself. Doing so might endanger the patient, damage the ventilator, and/or void your warranty. Only personnel authorised and qualified by Covidien should repair, open or service the ventilator. Replacing the Air Inlet Filter WARNING Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator.
  • Page 102: Recommended Schedule Of Maintenance

    For all additional accessories not necessarily considered as consumables consult the manufacturer’s recommendations. To prevent any risk of cross contamination Covidien recommends the use of STERIVENT filters (Ref: 351/5856 or equivalent) to protect the patient outlet port. WARNING Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator.
  • Page 103: Replacement Of The Internal Battery

    3, “Alarms and Troubleshooting”. If you cannot determine the cause of the problem, contact your equipment supplier or Covidien. For more information and local Covidien Technical Service Contact details, refer to chapter , “Technical Support” in Preface chapter. User’s Manual...
  • Page 104 Routine Maintenance This page intentionally blank User’s Manual...
  • Page 105: A Specifications

    Physical A Specifications Physical Table A-1. Physical Description (Excluding Accessories) Ventilator Weight 9.9 lb. (4.5 kg) Ventilator Dimensions 9.25 in wide x 12.40 in deep x 6.0 in high (235 mm wide x 315 mm deep x 154 mm high) Connectors Inspiratory limb connector: ISO 22 mm (OD) conical Oxygen inlet: Female Connector with valve...
  • Page 106: Indicators And Alarms

    Specifications Table A-3. Internal Lithium Ion Battery (Continued) Average operating time at 25 °C (± 5 ºC) with a fully charged battery (having less than 50 charge/discharge cycles) at the following displayed values: Vt = 200 ml (± 5 ml), PIP = 10 mbar (± 2 mbar), Rtot = 20 bpm 5 hr.
  • Page 107: Performance

    Performance Table A-6. Alarm Indicators High Priority Medium Priority Red flashing LED Yellow flashing LED Table A-7. Audio Alarms Audio Paused Alarm Volume 60 s ± 1 s 65 to 85 dB (A) ± 10% at 1 meter Performance A.4.1 Specifications Table A-8.
  • Page 108: Range, Resolution, And Accuracy

    Specifications Range, Resolution, and Accuracy Table A-10 lists the ranges, resolutions, and accuracies for ventilator settings, alarm settings, and patient data. Table A-10. Ventilator Range, Resolution, and Accuracy Ventilator Settings Range, Resolution, and Accuracy Mode Range: P A/C, PSV, CPAP Resolution: N/A Accuracy: N/A Default value: P A/C...
  • Page 109 Range, Resolution, and Accuracy Table A-10. Ventilator Range, Resolution, and Accuracy (Continued) Ventilator Settings Range, Resolution, and Accuracy Inspiratory sensitivity (I Sens) Range: 1P-5 Resolution: 1 Accuracy: NA Default value: 2 in CPAP, I Sens is set to 2 and is not adjustable Exhalation sensitivity (E Sens) Range: 5% to 95% of peak flow Resolution: 5%...
  • Page 110: Environmental

    Specifications Table A-10. Ventilator Range, Resolution, and Accuracy (Continued) Ventilator Settings Range, Resolution, and Accuracy Minimum Inspiratory Time (Min I time) Range: 0.1 to 2.8s Resolution: 0.1 s Default value: AUTO (Rise time + 300 ms) Depends on: Max I Time, Backup R, Rise time Maximum Inspiratory Time (Max I time) Range: 0.8 to 3 s Resolution: 0.1 s...
  • Page 111: Pneumatic

    The use of any accessory other than those specified, with the exception of the power supplies or cables sold by Covidien, may lead to an increase in electromagnetic emissions or a decrease in the equipment protection against electromagnetic emissions. If the ventilator is used adjacent to such accessories or stacked with such devices, the ventilator’s performance should be monitored to verify normal operation.
  • Page 112: Table A-19. Electromagnetic Emissions

    Specifications Table A-19. Electromagnetic Emissions The Puritan Bennett™ 520 Ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment. RF emissions Group 1 The ventilator uses RF energy only for its internal functions.
  • Page 113: Table A-20. Electromagnetic Immunity

    Manufacturer’s Declaration Table A-20. Electromagnetic Immunity The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should ensure that it is used in such an environment. Electromagnetic Immunity Test IEC / EN 60601 Test Level Compliance Level Environment–Guidance Electrostatic discharge...
  • Page 114: Table A-21. Electromagnetic Immunity - Conducted And Radiated Rf

    Specifications Table A-21. Electromagnetic Immunity – Conducted and Radiated RF The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment. IEC / EN 60601-1-2 Test Electromagnetic Environment–...
  • Page 115: Table A-22. Recommended Separation Distances

    Manufacturer’s Declaration Table A-21. Electromagnetic Immunity – Conducted and Radiated RF (Continued) Note: • At 80 MHz and 800 MHz, the higher frequency range applies. • These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
  • Page 116: Table A-23. Compliant Cables And Accessories

    Specifications Table A-23. Compliant Cables and Accessories Cable or Accessory Maximum length UK AC power cable assembly 1.8 m (5.9 ft) Japan AC power cable assembly 1.8 m (5.9 ft) China AC power cable assembly 1.8 m (5.9 ft) South Africa AC power cable assembly 1.8 m (5.9 ft) India AC power cable assembly 1.8 m (5.9 ft)
  • Page 117 Standards Compliance and IEC Classification Particular Standards • Lung Ventilator for Medical Use-Particular Requirements for Basic Safety and Essential Performance Part 6: Home-care ventilatory support devices EN ISO 10651-6: 2009. • Lung Ventilators for Medical Use- Particular Requirements for Basic Safety and Essential Performance Part 1: Home Care Ventilator Support YY 0600.1-2007 (ISO 10651-6:2004, MOD).
  • Page 118 Specifications This page intentionally blank A-14 User’s Manual...
  • Page 119: B Operational Verification Checklist

    If the ventilator fails any of the safety checks below, or if you cannot complete these checks, refer to section 3.8, “Troubleshooting, ” on page 3-14 or call the equipment supplier or Covidien (refer to section 8.3, “Service Assistance, ” on page 8-3).
  • Page 120 Operational Verification Checklist This page intentionally blank User’s Manual...
  • Page 121: C Alarms Tests

    (refer to chapter 3, “Alarms and Troubleshooting”) of this manual or call your equipment supplier or Covidien (refer to section 8.3, “Service Assistance,” on page 8-3 Note: Many of the functions of the ventilator are not accessible when the Locking key is enabled.
  • Page 122: Occlusion Test

    Alarms Tests • the “AC POWER DISCONNECTION“ alarm activates • an audible alarm sounds • the DC POWER indicator illuminates if the DC power source is connected; otherwise, the INTERNAL BATTERY indicator illuminates 2. Press the ALARM CONTROL key twice to reset the alarm. Reconnect the ventilator to its AC power supply.
  • Page 123: Involuntary Stop Test

    Involuntary Stop Test 1. Disconnect the AC power supply cable and the DC power cable (if it is connected) from the ventilator. • a POWER DISCONNECTION alarm will trigger. 2. Press the ALARM CONTROL key twice to pause the alarm. Ensure that the following events occur: •...
  • Page 124 Alarms Tests This page intentionally blank User’s Manual...
  • Page 125: D Unpacking And Preparation

    Never use a ventilator or any components or accessories that appear to be damaged. If any signs of damage are evident, contact your equipment supplier or Covidien. 4. Clean the ventilator with a mild soap solution, if necessary (refer to chapter “Cleaning”).
  • Page 126: Figure D-1. Puritan Bennettâ„¢ 520 Ventilator

    Unpacking and Preparation Figure D-1. Puritan Bennett™ 520 Ventilator Figure D-2. Dual Bag User’s Manual...
  • Page 127: E Parts And Accessories

    Table E-1 provides a list of accessories that are available for the Puritan Bennett™ 520 Ventilator. To order parts or accessories, contact your equipment supplier or Covidien representative. Note: The ventilator is delivered with the following items: a printed User’s Manual, a CD with Clinician’s Manual (printed copy available upon request);...
  • Page 128 Table E-2 provides a list of consumable parts available for the ventilator. WARNING To ensure proper performance of the ventilator, use a patient circuit recommended by Covidien in this manual; refer to chapter 4, “Installation and Assembly” Appendix E, “Parts and Accessories”.
  • Page 129: F Glossary

    F Glossary AC Power Alternating current. Alarm Pause The audible and visual alarms cease and the symbol appears. The symbol will remain until the cause of the alarm is addressed. For example, when the ventilator is running on internal battery, the AC Disconnection alarm may be paused, and the alarm paused symbol will appear until the device is plugged into AC.
  • Page 130 Glossary Bias flow Turbine flow during exhalation phase through the patient circuit to avoid rebreathing. An abbreviation for “breaths per minute, ” which is the unit of measure for breath rate (see below). Breath Rate (Back Up R) The total number of breaths, both machine and spontaneous, delivered by a ventilator in one minute. Caregiver An individual who assists a patient with the tasks of daily living.
  • Page 131 Inspiratory Sensitivity (I Sens) Level of inspiratory effort the patient has to provide during the initiation of a machine breath. The sensitivity levels (from 1P to 5) correspond to differences in flow compared to the bias flow. Level 1P is the most sensitive (for a pediatric use) and requires the least effort to trigger a breath.
  • Page 132 Glossary Mbar An abbreviation for “millibar“ which is a unit of measure for atmospheric pressure. Mean Airway Pressure Average patient pressure during each breath. Minimum Exhalation Time Minimum exhalation time before allowing the patient inspiratory trigger. Minimum Inspiratory Time Minimum inspiratory time before allowing the patient to exhale. M Vol (Minute Volume) Flow delivered at each breath to the patient is measured by the inspiratory flow sensor and that measurement is used to calculate minute volume (Vt x Rtot)
  • Page 133 Pressure Support (P Support) Augmentation of the patient’s ventilation synchronously with inspiratory effort until a preset pressure is met. Pressure is maintained until inspiratory flow is reduced to a percentage of peak flow that depends on the exhalation sensitivity setting for the inspiration, when the ventilator cycles into exhalation.
  • Page 134 Glossary Vent Time (Ventilation Time) The ventilation duration data is based on the patient counter and shows the total ventilation time in hours and minutes over the previous 24 hour period. Vt Target (Target volume) The Vt Target feature enables the ventilator to reach a set volume of gas to be delivered to the patient using a range of pressures in the pressure-based ventilation modes.
  • Page 135 Index DEVICE FAULT3 ......3-10, 3-16 DEVICE FAULT5 ......3-10, 3-16 AC power DEVICE FAULT7 .
  • Page 136 Index APNEA alarm message ..... .3-8, 3-14 COOLING FAN alarm message... . . 3-10, 3-16 Audible alarms silencing .
  • Page 137: Maintenance

    Index NO PROXIMAL LINE2 alarm message ..3-12, 3-19 Notes I/O switch (figure)....... . 5-2 definition of .
  • Page 138 Index maintenance ....... . 1-11 internal battery ......6-4, C-2 oxygen .
  • Page 139 Index display of ........5-2 skipping .
  • Page 140 This page intentionally blank...
  • Page 142 10056025 Rev C 03/2012 COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and/or internationally registered trademarks of Covidien ™ Trademarks of a Covidien company. ™* Trademarks of their respective owner. © 2012 Covidien. Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.

Table of Contents