Specifications; Sequencer Specifications; Timebase Specifications; Stimulus Timing Generator Specifications - Agilent Technologies VXI 75000 C Series Hardware Manual

160mhz timing module
Table of Contents

Advertisement

Agilent E1450 160 MHz Timing Module

Sequencer Specifications

Memory Depth: 65,536 (64k) vectors. Multiple test sequences may co-reside in memory.
Functions: Timing Cycle Selection, Stopping Point, Breakpoints, Markers, and Trigger Arming, all
programmable on a cycle-by-cycle basis.

Timebase Specifications

Timing Cycle Duration:
Minimum: 50ns or 4 Subcycles, whichever is greater
Maximum: 1024 Subcycles (see NUMBER OF TIMING CYCLES below)
Resolution: One Subcycle
Subcycle Period:
Minimum: 6.25ns
µ
Maximum: 409.6
s
Resolution: 6.25ns
Accuracy: 0.01%
Number Of Timing Cycles: Up to 256. The total length of all Timing Cycles may not exceed 1024
Subcycles.

Stimulus Timing Generator Specifications

Function: Provides six Stimulus Pattern Clock signals which define edge placements for E1451A/52A
Pattern I/O Ports configured in Stimulus mode. (Each I/O Port is programmed to use one Pattern Clock
as its timing reference.) Pattern Clocks are independent of each other and may have different delays in
different Timing Cycles.
Stimulus Pattern Clock Delay (from beginning of Timing Cycle)
Minimum: 0 Subcycles
Maximum: Timing Cycle Duration - 1 Subcycle
Resolution: 1 Subcycle
NOTE: No sequence of Timing Cycles may be allowed to cause
Pattern Clocks to occur less than 50ns apart.