1
Introduction
Specifications
Frequency Range and Attenuation
Table 2
Frequency Range and Attenuation
Product
Frequency Range
Attenuation
Steps
Attenuation Accuracy
Table 3
Attenuation Accuracy
8494A/B 8495A/B
8496A/B
Attenuation
Selection (dB)
1
10
2
20
3
30
4
40
5
50
6
60
7
70
8
80
9
90
10
100
11
110
12
8494A
8494B
dc to 4 GHz
dc to 18 GHz
0 dB to 11 dB
0 dB to 11 dB
1 dB
1 dB
(±dB): (Referenced from 0 dB)
8494A
8494B
dc–4
dc–12.4
12.4–18
GHz
GHz
GHz
0.2
0.3
0.7
0.2
0.3
0.7
0.3
0.4
0.7
0.3
0.4
0.7
0.3
0.5
0.7
0.3
0.5
0.8
0.4
0.6
0.8
0.4
0.6
0.8
0.4
0.6
0.8
0.4
0.6
0.9
0.5
0.7
0.9
8495A
8495B
dc to 4 GHz
dc to 18 GHz
0 dB to 70 dB
0 dB to 70 dB
10 dB
10 dB
8495A
8495B
dc–4
dc–12.4
12.4–18
GHz
GHz
0.2
0.5
0.4
0.7
0.5
0.9
0.7
1.2
0.8
1.5
1.0
1.8
1.2
2.1
–
–
–
–
–
–
–
–
Agilent 8494/95/96A/B Attenuators Operating and Service Manual
8496A
dc to 4 GHz
0 dB to 110 dB 0 dB to 110 dB
10 dB
8496A
dc–4
dc–12.4
GHz
GHz
GHz
0.6
0.2
0.5
0.8
0.4
0.7
1.2
0.5
0.9
1.6
0.7
1.2
2.0
0.8
1.5
2.4
1.0
1.8
2.8
1.2
2.1
–
1.3
2.4
–
1.5
2.7
–
1.6
3.0
–
1.8
3.3
8496B
dc to 18 GHz
10 dB
8496B
12.4–18
GHz
0.6
0.8
1.2
1.6
2.0
2.4
2.8
3.2
3.6
4.0
4.4