Agilent E1453 Timing Pod Description

The optional E1453 Timing Pod (Figure 1-3) extends measurement accuracy to a DUT located up to two
meters from the front panel of the Model D20. The pods are active devices which not only improve the
Model D20's ability to drive DUT inputs, but also minimize loading on DUT outputs. Signal delay in the
pod and cable is automatically compensated for by the Timing Module when power is applied.
The timing pod buffers the eight control signals out to the DUT as well as the 10 "Q" lines, a ready line
and a trigger line into the E1450 module. When the E1453 is not used, the Control and "Q" lines can be
accessed via the E1450's pod connector; trigger and ready lines are available on SMB connectors. The
pod is supplied with a 2.1-meter cable.
Note
More detailed information about each of the inputs and outputs of the Timing Module is contained in
"Specifications", Chapter 3.
E1450 Timing Module
E1453 Timing Pod
Hardware Descriptions 1-3