160mhz timing module
Table of Contents

Advertisement

Address Space
Outside the command module
Inside the command module
Important
These modules are D16-only devices. That is, they may only be accessed via 16-bit words on even addresses.
Attempts to access these modules via 8-bit bytes (D8) or 32-bit words (D32) will result in a bus error.
3-4 Register Maps
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Hexadecimal Register Address
OFFSET + E200h + 22h =
OFFSET + E222h
1FE200h + 22h = 1FE222h
Decimal Register Address
OFFSET + 57,856 + 34 =
OFFSET + 57,890 + 34 = 52,258
2,089,472 + 34 = 2,089,506