Recommended Fixturing Techniques

The pod mating connector (DUT side) is a 50-pin (2 x 25) male dual in-line connector with
0.025-inch (0.64mm) round or square pins on 0.100-inch (2.54mm) centers. This connector is not
supplied with the product but is available from electronic supply houses.
Do not connect the device under test (DUT) directly to the pod connector. Whenever possible, use
intermediate wiring between the pod mating connector and the test fixture connector (such as the
PC Board Edge Connector shown below). This minimizes the number of insertions/removals made
to the pod connector which extends its lifetime.
Keep all wiring to the DUT as short as possible.
CAUTION
Clearly label the wiring and connectors to identify timing connections and I/O connections. Equipment
or DUT damage can occur if these connections are accidentally reversed.
2-8 Configuration and Wiring
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 2-7. Typical Pod to DUT Wiring