Resuscitation Module

1. To use the Suction feature, turn it on using the On/Off
switch.
NOTE:
The gas supply must be turned on.
2. To adjust suction from 0 kPa to -19.99 kPa, use the
Suction control knob.
3. Before use, verify that the suction gauge reads -19.99
kPa (-150 mmHg) ± 0.53 kPa (4 mmHg) when suction
control is set to maximum and the suction output is
occluded. The suction output may be occluded at the
suction jar.
4. Verify that the suction switch turns the suction
completely OFF and completely ON.
WARNING:
There are potential hazards associated with the delivery of supplemental oxygen. If it is
necessary to administer oxygen, notify the attending physician immediately. Failure to do so
could result in patient injury.
WARNING:
Always monitor the airway pressure, and set appropriate pressure relief during infant
resuscitation. Failure to do so could result in patient injury.
WARNING:
For prolonged resuscitation, use a heat and moisture exchanger or heated humidification.
Failure to do so could result in patient injury.
5. On units with the AutoBreath™ Infant Resuscitator, perform the following:
a. To turn the timing circuit on and off, use the On/Off switch.
b. To adjust the Positive End Expiratory Pressure (PEEP) in the patient circuit, use the PEEP
Min/Max control.
c. Adjust the inspiratory and expiratory times using the Rate Control.
The actual display may differ in appearance or configuration.
5 - 11