Suction

A gas-powered, venturi-generated vacuum drives the
suction circuit, which provides a negative pressure for
suctioning the patient airway. This suction device is
specifically intended for removal of mucus from neonates
post delivery. The suction negative pressure is indicated on
the suction gauge. Adjust suction with the Suction control,
and switch it on or off with the On/Off switch. A fixed
relief valve limits the maximum suction pressure to 150 mmHg (80 cm H
O/mbar). This suction module
2
contains an On/Off switch and a suction control knob. The suction control knob provides suction
adjustment between 0 to 19.99 kPa (0 to 2.90 psi), minimum of 16 LPM at maximum.
NOTE:
The disposable suction bottle has a built-in filter.
Disposable suction bottle, 800 mL (MU10918) and the
associated filter (MU10915) (WBR82 only)
Reusable suction bottle, 600 mL (MU22310) and the
associated filter and tubing (MU10923) (RW82 only)
Micro-volume suction bottle, 40 mL
(MU10913) and the associated filter
(MU11147), or box of 10 disposable micro-volume suction
bottles, 40 mL (MU10917), or micro-volume suction kit
(MU10916)