Maintenance - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Maintenance

WARNING:
Risk of electric shock. Service personnel could be injured by electric shock. Before servicing
work is carried out, disconnect all electrical and gas connections from voltage and gas supply
systems.
WARNING:
Risk of injury or equipment damage. Preventive maintenance performed by unauthorized
personnel could result in personal injury or equipment damage. Only facility-authorized service
personnel must perform preventive maintenance on the Resuscitaire Radiant Warmer products.
WARNING:
Risk of patient injury or equipment damage. Loose particles from the heater element could
cause patient injury or equipment damage. During maintenance, inspect the secondary reflector
directly under the warmer heater element for particles. If particles are present, replace the
heater element.
For disposal of consumable materials, see "Disposal" on page 6-10.