Assembly Of The Resuscitaire Birthing Room Warmer (Wbr82); Warmer - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Assembly of the Resuscitaire Birthing Room Warmer (WBR82)

CAUTION:
When removing the equipment from the carton, use care not to scratch or otherwise damage
unprotected surfaces.

Warmer

1. If a monitor shelf is desired on the unit, install the
weights provided with the monitor shelf in the bottom
of the lower post. Refer to the instructions provided
with the monitor shelf.
2. Remove the two #8-32 x ¾" screws (A) that secure the
controller (B) to the head support plate (C). Retain the
two #8-32 x ¾" screws.
3. Slide the controller (B) out through the front of the
upper post (D).
WARNING:
Use two persons of sufficient strength to lift the
upper post. Failure to do so could result in
personal injury or equipment damage.
4. Using two persons of sufficient strength, lift the upper
post (D), and rest it on top of the opening of the base
weldment (E).
5. Fully extend the suction hoses (F) from the upper post
and the base weldment, and then connect them to each
other.
6. Carefully feed excess hose into the upper post (D).
7. Release the suction hoses from the Neat-clip located
inside the base weldment.
8. If necessary, reposition the upper post on the base
weldment assembly.
9. Install the four #10-32 x 3/8" Nylok screws (G)
provided to secure the upper post to the bassinet/cart
assembly.
CAUTION:
Do not kink the suction hoses. Equipment
damage could occur.
10. Being careful not to kink the suction hoses, push them into the upper post.
4 - 2
B
C
A
D
D
F
E
G