Controller

SHOCK HAZARD:
Plug the power cable only into a properly grounded wall receptacle that is approved for hospital
use and of the correct voltage. Do not use extension cords or an AC receptacle box for this
device. Personal injury or equipment damage could occur.
WARNING:
The warmer is intended for use with infants in the cart and bassinet. If the warmer is used to
warm the infant held by the mother immediately after delivery, follow all warnings and
instructions. Pay particular attention to heater-to-mattress distance and skin probe location.
Avoid placing objects near the mother that can absorb heat and become hot to the touch.
Personal injury could occur.
WARNING:
Radiant warming increases insensible water loss. Consider appropriate measures to maintain
fluid balance. Failure to do so could result in patient injury.
1. Plug the unit into the appropriate AC line power source.
2. Switch on the circuit breaker on the rear panel.
3. Switch on the On/Off switch on the front panel.
4. Observe the functional test.

Pre-Warm Mode

NOTE:
During Pre-Warm Mode, the alarms are deactivated.
NOTE:
Selecting Manual or Baby Mode and then returning to Pre-Warm Mode during the 3 min of 100 % power
or the 12 min of 60 % power automatically reduces the power to 30 %.
NOTE:
The patient end of the skin probe must be attached or a probe alarm occurs. This alarm indicates a
difference of 0.4 °C between the two thermistors in the skin probe.
NOTE:
On units with software version 1.08, if a skin probe is connected while the unit is in Pre-Warm Mode,
power reduces to 30 %.
After the function test is complete, Pre-Warm Mode activates. The Heater Power indicator is at 100%
(all lights on) for 3 min. This indication reduces to 60% (six lights on) for 12 min, and then reduces to
30% (three lights on).

Manual Mode

WARNING:
Never place objects, such as blankets, clothing, diapers, or sterile packs, on top of the warmer.
These objects can interfere with the proper cooling of the warmer head and can fall onto the
mattress. Patient injury or equipment damage could occur.
1. Using the Mode Select key, select Manual Mode. The Manual indicator illuminates. If the Manual
indicator does not illuminate, do not use the warmer in Manual Mode.
5 - 12