Alarms - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Alarms

Feature
High Temperature
Check Patient
APGAR Timer
Power Fail
Probe Fail
System Fail
Baby Temperature
Electrical Module, Audio Ramping
Alarms (80 dBa maximum for infant)
Blender Module, Pneumatic Audio
Alarm Type
Activates if the skin probe is attached and the skin probe
reaches 39.0°C ±0.2°C (102.2°F). Resets at 38.5°C (101.3°F).
Activates in Manual Mode after 10 min. Remains on with
acoustic alarm signal every 30 s for 5 min for a total of 15 min.
The heater then turns off.
Activates at the 1 min, 5 min, and 10-min APGAR timer inter-
vals.
Activates when there is a loss of power. Stays activated for
>10 min.
Activates if the skin probe fails (open or short).
Indicates a system failure. Refer the unit to service immediately.
Activates if the baby temperature fluctuates 1 °C above or
below the setpoint.
Tone frequency: 1.2 kHz maximum with a three-stage sound
level: 15 s low, 15-s medium, and then high.
Vibrating reed, pneumatically powered.
Specification
2 - 25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents